Chapter 4 Instrument Functionality
索引
4
3
2
1
7
6
5
10
9
8
The buzzer is configured when turning on the instrument.
Turning off the buzzer:
Turn on the instrument by pressing the
[POWER] key while holding down the [AVG]
key to disable the buzzer.
Re-enabling the buzzer:
Turn on the instrument by pressing the
[POWER] key while holding down the [AVG]
key again to re-enable the buzzer.
The display backlight remains on at all times and cannot be
turned off.
The Auto Power Save function is configured when turning on the
instrument.
• The function operates automatically when the instrument is
turned on.
• After five minutes of inactivity and no measurement signal
detection, the instrument turns off automatically. This function
is only operative when using battery power.
• "APS" starts to blink at 30 seconds before turning off.
Disabling Auto Power Save: :
Turn on the instrument by pressing the [POWER]
key while holding down the [HOLD] key to disable
Auto Power Save.
Re-enabling Auto Power Save:
Turn off the instrument and then turn it back on.
The APS function will be enabled.
Configuring the Buzzer
Display Backlight
Auto Power Save (APS) Function
Buzzer disabled
Buzzer enabled