3.2 General Specifications
Function specifications (underline: default value)
Data hold function
Backlight functionAutomatically turns off approx. 2 min. after last key operation.
Filter function
Resistance
measurement
filter function
Moving average time: Max. 9 sec.
Current
measurement
filter function
Cutoff frequency: 180 Hz ±30 Hz (-3 dB)
Alarm function
Resistance alarm
function
Resistance measurement mode alarm: Beeps when measured
value is less than or greater than threshold.
Current alarm
function
Current measurement mode alarm: Beeps when measured value
is less than or greater than threshold.
Alarm Hi/Lo Separate Hi/Lo settings for resistance measurement and current
measurement
Resistance measurement: Hi.AL/Lo.AL
Current measurement: Hi.AL/Lo.AL
Alarm threshold
setting range
Resistance measurement: 0.02 Ω to 1,600 Ω
Resistance measurement initial value: 25.0 Ω
Current measurement: 0.05 mA to 200.0 mA, 0.201 A to 60.0 A
Current measurement initial value: 1.00 mA
Memory function
Memory capacity 2,000 values
Auto-power-save
function
Instrument automatically turns off approx. 5 min. after last key
operation.
Function mode
Moving between
items
After selecting setting item with ▼ and ▲ keys, accept with OK
key.
Resistance alarm
function
AL Ω: Resistance Hi/Lo, threshold settings
Current alarm
function
AL A: Current Hi/Lo, threshold settings
Loading memory
values
MEM READ: Load memory values.
Clearing
memory values
MEM CLR: Clear last saved memory value or all values.
Bluetooth
®
operating setting
BT: ON/OFF (FT6381)