31
Comparator Function
3.4 Comparator Function
This function compares the measured value with the preset value and judges
whether the result is PASS (good) or FAIL (defective).
Comparator settings for each range will be saved, even if the rotary switch is
turned off.
See the table on the next page for criteria that can be set.
Indication lights up
PASS (good) judgment FAIL (defective) judgment
LED
display
Backlight:
No change
Lights up in
green*
Backlight: Lights
up in red
Lights up in
red*
* When using the L9788-10 Test Lead with Remote Switch (Red)
Type of measurements that can be judged
Function
PASS judgment FAIL judgment
State of
measured
value
Buzzer Backlight Buzzer
Insulation
Resistance
Criterion or
higher
Short beep
Lights up in
red
Long beep
Low resistance
Criterion or
lower
Long beep Short beep
PV
Ω
Criterion or
higher
Short beep Long beep
Voltage Comparator cannot be set.