PVΩ Measurement Function (Model IR4053 and IR4055 only)
WARNING
Observe the following to avoid electric shock, short
circuit, or damage to the instrument.
Turn off any disconnector devices and separate
from the power conditioner before starting the
measurements for the solar battery panel.
Do not attempt to measure insulation resistance
on a live conductor. Doing so could damage the
instrument or cause an accident that might result
in injury or death. Always turn off power to the
measuring object before starting.
Photovoltaic cells produce electricity continuously
during daylight hours, resulting in hazardous
voltages. Exercise care during measurement to
avoid electric shock.
Do not touch any metal parts such as connection
box and disconnector devices directly with bare
hands. Doing so may cause electric shocks due to
the voltage of the generator.
Maximum rated voltage between terminals of the
IR4053 and IR4055 is 1000 V DC/600 V AC. Do not
use the instrument for equipment with rated voltage
over 1000 V DC or 600 V AC. Doing so may cause
electric shock or failure.
If the solar battery panel has failed, do not perform
any insulation resistant measurement. Doing so
may damage the bypass diode of the solar battery
panel.
1
2
4
5
6
7