Section 2 – Installation
SECTION 2 – INSTALLATION
GENERAL
This section provides detailed installation instructions for the TLi
IQ
System. Follow installation steps carefully to insure
proper installation and operation.
ENVIRONMENTAL FACTORS
The TLi
IQ
System has been designed to be safe under the following conditions: Indoor use; Altitudes up to 2000m;
Maximum relative humidity of 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C; Main
supply voltage fluctuations not to exceed ± 10% of the nominal; transient overvoltages according to Installation Category
and Pollution Degree 2. However, as with all electronic instruments, prolonged exposure to high temperature and humidity
should be avoided. The operating temperature should be held relatively constant. The optimum operating temperature is
18° to 30°C (64° to 86°F). Before operating, allow the instrument to equilibrate to room temperature.
Place the instrument where it will not be subjected to extreme temperature variations (e.g., near open windows, ovens,
hot plates, radiators, direct sunlight).