TLi
IQ
System User Manual 3 2
Section 3 General Operating/Testing Instructions
SET CALIBRATION
NOTE: Calibration must be set when changing cassette lots.
1. From the Main Menu, press 2 to select SET CALIBRATION.
2. Enter USER ID and press ENTER.
3. Enter the CASSETTE LOT# (on cassette pouch) and press ENTER.
The lot number must be entered to proceed to the next step.
fFN MAIN MENU
1-TEST PATIENT
2-SET CALIBRATION
3-DAILY QC
SET CALIBRATION
USER ID
>JOHN SMITH
ENTER-CONFIRM
SET CALIBRATION
CASSETTE LOT#
>L1002
ENTER-ACCEPT