System User Manual 3 − 7
Section 3 – General Operating/Testing Instructions
DAILY QC
NOTE: TLi
IQ
QCette SETUP must be performed PRIOR to running the
QCette as a quality control device. See Section 4 for TLi
IQ
QCette
SETUP.
1. From the Main Menu, press 3 to select DAILY QC.
2. Enter USER ID and press ENTER.
3. Enter the QCette SN (on QCette plastic housing) and press ENTER.
The serial number must be entered to proceed to the next step. The
correct format is 6 numbers (e.g., 001004). Enter all leading zeros.
NOTE: The serial number entered at daily QC must be identical to the
serial number entered at TLi
IQ
QCette setup.