Section 3 General Operating/Testing Instructions
VIEWING RESULTS ON SCREEN
Upon completion of every test, the analyzer will automatically display the
results on up to three screens. Each screen can be accessed by using
the and keys. To print the result record, press the ENTER/PRINT
key from any screen.
NOTE: Internal controls are performed automatically during each Rapid
fFN test. These internal controls check for (1) a threshold level of
signal at the procedural control line, (2) proper sample flow
across the Rapid fFN Cassette, (3) absence of conjugate
aggregation, and (4) proper functioning of the TLi
IQ
analyzer
hardware.
FETAL FIBRONECTIN
03:00 PM 10/26/07
PT:JANE DOE 123
RESULT:POSITIVE
CAL CODE:FG56
CASSETTE LN:L1002
ANALYZER ID:00426
ESC-MAIN MENU
INTERNAL CONTROLS
USER:JOHN SMITH
ANALYZER: PASS
CASSETTE: PASS