TLi
IQ
System User Manual 4 4
Section 4 Software Functions - Detailed Descriptions
The CASSETTE LOT# must be entered to proceed to the next step. The
CASSETTE LOT# is located on the cassette pouch. The software
requires that the lot number is entered in the correct format: one alpha
character followed by four numeric characters (e.g., L1002).
The CALIBRATION CODE# must be entered to proceed to the next step.
The CALIBRATION CODE# is located on the cassette box. The software
requires that the code number is entered in the correct format: two alpha
characters followed by two numeric characters (e.g., FG56).
NOTE: The calibration code is established by Hologic for each lot of
Rapid fFN Cassettes.
Calibration Data Record - This record is displayed over two screens.
Each screen can be accessed by using the and keys. The
complete record will be printed automatically if AUTOPRINT is set to ON,
or it may be printed/reprinted by pressing the PRINT/ENTER key.
SET CALIBRATION
CALIBRATION CODE#
>FG56
ENTER-ACCEPT
FETAL FIBRONECTIN
03:00 PM 10/26/07
SYSTEM CALIBRATED
USER:JOHN SMITH
CAL CODE:FG56
CASSETTE LN:L1002
ANALYZER ID:00426
ESC-MAIN MENU
SET CALIBRATION
CASSETTE LOT#
>L1002
ENTER-ACCEPT