TLi
IQ
System User Manual 4 13
Section 4 Software Functions - Detailed Descriptions
TEST WARNING: This message will be displayed if ESC was pressed
during testing. Lines 1 and 2 will flash prompting the user to select
ENTER to continue test, or ESC to end test. This message will hold for
5 seconds and then revert back to its respective screen.
QCette Data Record This record will be displayed on two screens.
Each screen can be accessed by using the and keys. The
complete record will be printed automatically if AUTOPRINT is set to ON,
or it may be printed/reprinted by pressing the PRINT/ENTER key.
QCette results are SYSTEM PASS, SYSTEM FAIL, or INVALID.
A FAIL or INVALID result should be repeated. (See Section 7, Items 9
and 10.)
Invalid results will not be stored in memory.
ARE YOU SURE YOU
WANT TO CANCEL TEST
ENTER-CONTINUE
ESC-MAIN MENU
QCette
03:00 PM 10/26/07
SYSTEM: PASS
USER:JOHN SMITH
CAL CODE:FG56
QCette SN:00104
ANALYZER ID:00426
ESC-MAIN MENU