TLi
IQ
System User Manual 4 17
Section 4 Software Functions - Detailed Descriptions
A two-minute timer starts during which time this message flashes and the
analyzer beeps. Add control sample and immediately press ENTER.
If the sample is not added and ENTER not pressed within allotted time,
the test process cannot continue. This message will be displayed. Press
ESC to return to the Main Menu. No record of the test will be held in
memory.
Once the sample is added, the analyzer will begin a 20-minute
incubation countdown. To abort the test, press ESC. Pressing ESC will
terminate the test and the data will be lost.
Upon completion of the incubation period, the analyzer will begin the
analysis of the cassette. Do not disturb the analyzer until the results are
displayed. The analysis will take approximately 2-3 minutes.
TEST IN PROCESS
DO NOT REMOVE
CASSETTE
19 MIN 56 SEC
TEST IN PROCESS
ANALYZING
DO NOT REMOVE
CASSETTE
ADD SAMPLE
AND IMMEDIATELY
PRESS ENTER
SAMPLE NOT ADDED
WITHIN ALLOTTED TIME
ESC-MAIN MENU