TLi
IQ
System User Manual 7 – 4
Section 7 – Troubleshooting
Ensure QCette is clean and not damaged. Repeat TLi
QCette setup as described in Section 4. If the QCette
setup fails a second time, call technical service.
Turn analyzer off and back on to reinitialize the system.
Repeat TLi
IQ
QCette setup as described in Section 4. If
the QCette setup fails a second time, call technical
during daily quality
control.
Ensure QCette is clean and not damaged. Repeat TLi
QCette daily quality control as described in Section 4. If
the QCette testing fails a second time, call technical
service.
Turn analyzer off and back on to reinitialize the system.
Repeat TLi
IQ
QCette daily quality control as described in
Section 4. If the QCette testing fails a second time, call