50489/GTEX Chem IIC/1805 UTImeter Gtex Chem IIC
9.3. Tape replacement
THE REPLACEMENT OF THE TAPE DOES NOT REQUIRE TO RE-CALIBRATE THE TEMPERATURE.
Follow the different sequences as described below. The Figure 13-1 : general assembly, list of the main
spare parts can also help.
9.3.1. Disconnecting the tape from the
sensor
Follow the instructions of section 10.4 "Sensing
probe replacement".
9.3.2. Disconnecting the tape from the
electronic box
− Unscrew with the 2.5 Allen key the 2 screws
(A) of the battery holder and pull it out as
shown on Figure 10-4.
− Unscrew with the 2.5 Allen key the 4 screws
(B) of the display unit and pull it gently out as
shown on Figure 10-4.
− Disconnect the connecting plug (C) as shown
on Figure 10-5 and remove the display unit.
− Unscrew with the 2.5 Allen key the tape holder
(G) by removing the 2 screws (F) and the
grounding cable (D) as shown on Figure 10-5.
Do not loose the 2 remaining screws that
secure the reel axle.
Figure 10-4
(D) grounding cable of tape
(E) black oversleeve of tape
(G) tape holder
(H) tape
(C) connecting plug
to display unit
(I) securing screw
(F) securing screw
Figure 10-5