48  www.tanksystem.com
11.4 Ullage and/or Interface troubles
11.6
Visual inspection for damaged or
missing parts
General condition:  missing parts
Display unit:   5-key control pad, buzzer, front face, 
LED, screen 
Sensing probe:   sensors broken, smashed or damaged 
Tape:   check at least the rst 3 m; wires still 
insulated, no breaks, no kinks, ...
Mechanical parts:  check housing, lid, axle, storage tube, 
wipers of tape cleaner, window wiper
Symptom Origin Action Section
The buzzer does not beep when  
the unit is switched on
Buzzer switched off or Press on “-” to reactivate it 9.6
Key-pad defective or
Pressing on “+” has no action  
Change the display unit
10.7
Buzzer defective
Press on “+”: “Settings” is displayed 
Change the display unit
10.7
The buzzer tones continuously  
when the sensing probe is  
in air or liquid or water
Battery too low Change the battery 10.3
The buzzer gives the water  
signal whatever liquid is gauged
Sensing head  
contaminated by  
conductive residues
Wash, clean and brush (soft brush) the 
sensing head or 
change the sensor
--- 
10.5
The buzzer gives the oil 
signal in water
Sensing head  
contaminated by  
non conductive residues
Wash, clean and brush (soft brush) the 
sensing head or 
change the sensor
--- 
10.5
11.5 Temperature troubles
Symptom Origin Action Section
“> 90°C” or “> 194°F” is 
displayed
Temperature to high
The temperature range shall be  
< 90°C / 194 °F
---
“< -40°C” or “< -40°F” is  
displayed
Temperature too low
The temperature range shall be  
> -40 °C/F
---
Temperature does not 
stabilise
Heated viscous liquid 
(such as heavy crude  
oils)
Check the stability in cold and hot water;  
if it is OK the problem is with the gauged  
liquid and not with the probe
---
Contaminated sensing 
probe
Clean the temperature electrode; remove  
any residues or sludge; check the stability  
in cold and hot water 
---