EasyManuals Logo

Honeywell Searchline Excel User Manual

Honeywell Searchline Excel
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56
MAN0530 Issue 10 - 11/09 Searchline Excel
2104M0506
TheproceduresforusingthesefunctionaltestltersonModstate7orhigherExcelunitsare
detailed as follows:-
1) Ensure the area is free from gas.
2) Connect the SHC-1 Handheld Interrogator to the unit to be tested.
3) Select DISPLAY from the Calibration menu.
4) For new units, refer to supplied test certificate for factory response to test filter
recommended.
5) Forotherunits,slideasuitablefunctionaltestlter(seetablesbelow)intothecowling
at the front of the receiver.
Receiver Gas Calibration Low Test Filter Response High Test Filter Response
Methane C or D 0.5 – 1.1 LEL.m F 1.5 – 2.8 LEL.m
Ethane C or D 0.6 – 1.3 LEL.m F 1.5 – 2.7 LEL.m
Propane B or C 0.5 – 1.2 LEL.m F 1.6 – 2.9 LEL.m
Butane B or C 0.6– 1.4 LEL.m E 1.4 – 2.6 LEL.m
Pentane A 0.6 – 1.1 LEL.m D 1.2 – 2.2 LEL.m
Hexane B or C 0.6 – 1.5 LEL.m E 1.4 – 2.7 LEL.m
Ethylene B or C 0.5 – 1.3 LEL.m F 1.8 – 3.4 LEL.m
Propylene A 0.5 – 1.0 LEL.m D 1.0 – 1.9 LEL.m
Butadiene A 0.5 – 1.0 LEL.m D 1.0 – 1.9 LEL.m
Methanol D 0.5 – 0.9 LEL.m E 1.0 – 1.9 LEL.m
Ethanol B or C 0.6 – 1.4 LEL.m E 1.5 – 2.8 LEL.m
6) Ifthisisthersttimethattheunithasbeentestedwiththisparticulartestlter,record
the gas reading that is produced. This reading should be used to verify correct operation
in the future.
7) Ifthisisnotthersttimethattheunithasbeentestedwiththisparticularlter,notethe
gas reading it produces and compare it to the original value recorded for this unit when
testedwiththislter.Providedthatthereadingiswithin±0.5LEL.moftheoriginalvalue,
the unit is working correctly. If the unit is working correctly, proceed to step 15.
8) If the gas reading is not within the expected range it is possible that the unit needs
cleaning or re-zeroing.
9) Removethetestlterfromthecowling.
10) Using a damp, lint-free cloth, clean the windows of the transmitter and the receiver.
11) Using the SHC-1 in CALIBRATE menu, re-zero the unit.
12) Re-insertthechosentestlterintothecowlinginfrontofthereceiver.
13) Verifythatthereadingproducedbythetestlterisnowwithin±0.5LEL.mofthereading
originallyobtainedwiththislter.Ifthegasreadingisstillnotwithintheexpectedrange,
check the system alignment (see section 3 of the Excel manual, 2104M0506) and repeat
the test in step 7.
4. MAINTENANCE

Other manuals for Honeywell Searchline Excel

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Searchline Excel and is the answer not in the manual?

Honeywell Searchline Excel Specifications

General IconGeneral
BrandHoneywell
ModelSearchline Excel
CategoryOther
LanguageEnglish

Related product manuals