4. Put the magnet over the ‘’ switch and remove to enter the Calibration menu.
5. The display will show the current gas reading, ‘ZEro’ and the ‘
’ icon ashes.
6. When the zero gas reading is stable use ‘ to conrm zero calibration.
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display shows ‘ZEro FAIL’ and
returns to menu mode).
8. If using zero-air, turn it off. Zeroing is complete and saved.
9. The display shows ‘SPAn’ with ‘YES’ ashing.
10. If span calibration is required use ‘’ proceed to the next step. If span calibration is not required,
use ‘▲▼’ to select ‘No’ and ‘’ to return to menu mode.
11. The display shows the current calibration span gas concentration ashing. Use ‘▲▼’ to change the
calibration span gas concentration, and ‘’ when required span calibration level is set.
12. Span calibration point is displayed with ‘YES’ ashing. Use ‘’ to conrm or ‘▲▼’ to select ‘No’ and
return and enter a new span calibration point.
13. The display will show the current gas reading, ‘SPAN’ and the ‘
’ icon ashes.
14. Connect the regulator to the span gas cylinder.
15. Apply the span gas to the sensor using the calibration gas ow housing. The live gas reading is
displayed. When the reading is stable, use ‘’ to conrm span calibration.
16. If the sensor cell has been replaced the following display may be shown.