EasyManua.ls Logo

Honeywell Series 3000 MkII - Zeroing and Span Calibration of Hydrogen Sulfide Sensors

Honeywell Series 3000 MkII
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
SERIES 3000 MKII TECHNICAL MANUAL S3KMKIIMAN_MAN0878_Issue 6_01/2014
8.2.2 Zeroing and span calibration of Hydrogen Sulde sensors
Hydrogen Sulde sensors can be affected by extreme humidity changes. A sudden increase in ambient
humidity can result in a short-term positive drift in the instrument’s reading. A sudden decrease in ambient
humidity can result in a short-term negative drift in the instrument’s reading. These are most likely to
be noticed during calibration with dry or cylinder gas.
When calibrating Hydrogen Sulde cartridges the following should be taken into account while following
the procedure in 8.2.1:
1. To zero the sensor, use a compressed air cylinder of 20.9%Vol oxygen (not Nitrogen). Do not use
background air.
2. Apply the gas to the sensor for three minutes before using ‘ to conrm the zero calibration.
3. If a span calibration is to be performed, the span calibration gas should be applied to the sensor
immediately after the zeroing procedure. Do not allow the sensor to return to ambient air conditions
between steps 2 and 3.
4. Apply the span gas for two minutes before using ‘ to conrm the span calibration.

Table of Contents

Related product manuals