Revision 4  Foundation Fieldbus RMA803 Remote Indicator User's Guide  Page 3 
1.2  Features and Options 
The RMA803 Foundation Fieldbus Remote Indicator has the following features: 
 
Table 1: Feature and Options 
Feature/Option  Standard/Available Options 
Communication Protocols  Fieldbus and only Advanced Display 
Human-Machine Interface (HMI) 
Options (Advanced Display) 
Advanced Digital Display: 
  0, 90, 180, & 270 degree viewing position adjustments 
  Standard and custom measurement units available. 
  Up to eight display screens with 3 formats are possible (Large 
PV or PV with Bar Graph or PV with Trend Graph) 
  Configurable screen rotation timing (4 to 30 sec) 
  Ability to enable/disable screen rotation  
Three-button programming (optional) 
Advanced display languages:  
East Asian: EN, CH, JP 
Western: EN, GE, FR, IT, SP, RU, TU 
Approvals (See Appendix C for details.)  ATEX, CSA, FM, IECx, NEPSI 
Mounting Brackets  Pipe mounting and wall mounting brackets in carbon steel and 
316 stainless steel. 
# of Devices/ Segments 
Entity IS model: 6 devices/segment 
Schedule Entries 
46 (max) 
# of VCRs 
50 (max), 50-link objects 
Compliance Testing 
Tested according to ITK 6.1.1 
Software Download 
Class-3 of the Common Software Download procedure as per 
FF-883.