Xenon/Granit User Guide 241
IMAGING COMMANDS
The scanner is like a digital camera in the way it captures, manipulates, and trans-
fers images. The following commands allow you to alter the way the scanner per-
forms these functions.
Note: If you are using the scanner in a stand, you must set the In-Stand Sensor Mode to Off 
in order to take images (see In-Stand Sensor Mode on page 124).
Single-Use Basis
Imaging Commands with their modifiers send instructions to the scanner on a sin-
gle-use basis, and take effect for a single image capture. Once that capture is com-
plete, the scanner reverts to its imaging default settings. If you want to 
permanently change a setting, you must use the serial default commands (see 
Chapter 13). When the serial default command is used, that selection becomes the 
new, permanent setting for the scanner.
Command Syntax
Multiple modifiers and commands can be issued within one sequence. If additional 
modifiers are to be applied to the same command, just add the modifiers to that 
command. For example, to add 2 modifiers to the Image Snap command, such as 
setting the Imaging Style to 1P and the Wait for Trigger to 1T, you would enter 
IMGSNP1P1T. 
Note: After processing an image capture command (IMGSNP or IMGBOX), you must follow 
it with an IMGSHP command if you want to see it on your terminal.
To add a command to a sequence, each new command is separated with a semico-
lon. For example, to add the Image Ship command to the above sequence, you 
would enter IMGSNP1P1T;IMGSHP. 
The imaging commands are:
Image Snap - IMGSNP (page 242)