3-100
Performance Tests
PDH Transmit/Receive Jitter Amplitude Accuracy
8 Check that the Synthesizer amplitude waveform on the Oscilloscope is between
1.76V pk_pk and 2.84V pk_pk.
9 Check that the HP 37718A Demod Output amplitude waveform on the
Oscilloscope is between 707 and 833mV pk_pk.