Testing
Performance
The
procedures
in
this
section
‘test
the
instrument's
electrical
performance
using
Performance
Specifications
given
in
chapter
1
as
performance
standards.
Specifications
applicable
to
individual
tests
are
noted
at
the
test
for
reference.
Testing
Interval
The
performance
test
procedures
may
be
performed
for
incoming
inspection
of
the
instrument
and
should
be
performed
periodically
thereafter
to
ensure
and
maintain
peak
performance.
The
recommended
test
interval
is
yearly
or
every
2,000
hours
of
operation.
Amount
of
use,
environmental
conditions,
and
the
user’s
experience
concerning
need
for
testing
will
contribute
to
test
requirements.
Chapter
4,
"Calibrating
and
Adjusting,"
for
information
about
the
calibration
cycle.
Equipment
Required
A
complete
list
of
equipment
required
for
the
performance
tests
is
in
the
Recommended
Test
Equipment
table
in
chapter
1.
Equipment
required
for
individual
tests
is
listed
in
the
test.
Any
equipment
satisfying
the
critical
specifications
listed
may
be
substituted
for
the
recommended
model.
The
procedures
are
based
on
the
model
or
part
number
recommended.
Self-Test
Verification
To
verify
system
operation
with
high
confidence,
without
the
test
equipment
and
time
required
for
performance
tests,
perform
the
self-tests.
These
internal
tests
verify
many
functions
on
the
main
assembly.
The
functions
tested
are
the
six
separate
memories
and
six
other
system
functions.
To
start
the
self-tests,
press
Utility
,thenpress
selftest
menu.
A
message
is
displayed
with
the
instruction
to
remove
all
inputs
to
the
instrument.
Press
test
all,
which
starts
a
loop
which
runs
all
the
self-tests
in
succession.
During
execution
of
the
self-tests,
the
following
messages
are
displayed
as
each
self-test
is
completed:
PASSED
Display
RAM
PASSED
Acquisition
RAM
PASSED
System
RAM
PASSED
Logic
Trigger
PASSED
Non-Volatile
RAM
PASSED
Analog
Trigger
PASSED
Protected
Non-Volatile
RAM
PASSED
Timebase
PASSED
System
ROM
PASSED
Time
Tag
PASSED
HP-IB
PASSED
D/A
Converter
PASSED
A/D
Converter
If
one
of
the
self-tests
fails,
FAILED
is
displayed
rather
than
PASSED,
and
a
16-bit
diagnostic
code
is
displayed.
This
code
is
used
by
factory
service
personnel
when
troubleshooting
the
main
assembly.
Failure
of
a
self-test
indicates
a
failure
on
the
main
assembly
which
must
be
returned
to
the
factory
for
service.
For
more
troubleshooting
information,
refer
to
chapter
5,
"Troubleshooting."