Power
Supply
This
section
provides
an
adjustment
procedure
for
the
power
supply.
To
adjust
the
power
supply
The
power
supply
should
rarely
need
adjustment.
Perform
this
procedure
only
if
the
supply
has
been
checked
and
is
known
to
be
incorrect.
Equipment
Required
Equipment
Critical
Specifications
Recommended
Model/Part
Digital
Voltmeter
Accuracy
+0.05%
HP
34407A
Procedure
1
Disconnect
the
instrument
power
cord
and
remove
the
cover.
If
necessary,
refer
to
the
procedures
in
chapter
6,
"Replacing
Assemblies."
2
Refer
to
figure
for
testpoint
and
adjustment
locations.
Ground
+5.2V
Power
supply
test
point
test
point
adjustment
a
\
ABA
SHaea
Qoaan
QQQH)
54542017
Power
Supply
Adjustment
Location
3
Connect
the
common
lead
of
the
voltmeter
to
COM
test
point.
4
Connect
the
positive
lead
of
the
voltmeter
to
+5.20
V
test
point.
5
Connect
the
oscilloscope
power
cord
and
set
power
switch
to
ON.
6
If
the
voltmeter
does
not
read
between
5.180
V
and
5.220
V,
adjust
for
5.200
V.
44