Calibrating
and
Adjusting
This
chapter
provides
firmware
and
hardware
calibration
and
adjustment
procedures
for
the
HP
54520
and
54540-Series
oscilloscopes.
Primary
adjustment
groups
are
the
following:
e
Power
Supply
Adjustment
e
Main
Assembly
Calibration
and
Adjustment
e
CRT
Monitor
Assembly
Adjustment.
Equipment
Required
Equipment
required
for
adjustments
is
listed
in
table
1-1,
"Recommended
Test
Equipment",
in
chapter
1
of
this
manual.
Any
equipment
that
satisfies
the
critical
specification
listed
in
the
table
may
be
substituted
for
the
recommended
model.
Equipment
for
individual
procedures
is
listed
at
the
procedure.
Calibration
Interval
There
are
two
types
of
calibration
for
the
HP
54520
and
54540-Series
oscilloscopes.
The
firmware
calibration
is
the
self
cal
(self-calibration).
Self-calibration
should
be
done
every
year,
or
every
2,000
hours
of
operation,
whichever
comes
first.
The
hardware
calibration
consists
of
adjustment
of
the
power
supply,
high-frequency
pulse
response,
and
CRT
display.
These
adjustments
only
need
to
be
done
under
circumstances
set
by
certain
needs,
which
are
explained
in
other
areas
of
this
guide.
The
need
for
calibration
or
adjustment
will
also
depend
on
your
experience.
For
replacement
assemblies,
adjustments
are
set
at
the
factory
when
assemblies
are
tested.
However,
some
adjustment
may
be
necessary
after
an
assembly
has
been
put
into
the
instrument.
Usually
the
only
assembly
that
requires
adjustment
is
the
assembly
replaced.
Cal
RAM
Checksum
Error
If
power
is
applied
to
the
instrument
and
the
message
"cal
ram
checksum
error
re-cal
instrument"
is
displayed,
all
firmware
calibration
procedures
must
be
performed.
See
"To
calibrate
the
firmware"
in
this
chapter.
If
the
instrument
does
not
pass
the
firmware
calibration,
repair
is
necessary.
Operating
Hints
Some
knowledge
of
operating
the
HP
54520
and
54540-Series
oscilloscopes
is
helpful.
However,
procedures
are
written
so
that
little
experience
is
necessary.
The
following
hints
will
speed
progress
of
the
procedures.
When
using
many
averages,
it
often
takes
awhile
for
a
waveform
display
to
stabilize
after
a
change.
When
a
front
panel
control
on
the
oscilloscope
is
changed,
averaging
automatically
restarts.
When
the
input
signal
or
an
adjustment
is
changed,
the
4-2

Table of Contents