3:
Testing
Performance
To
test
time
measurement
accuracy
Press
Autoscale_,
then
set
the
following
parameters.
:
Menu
(hardkey)
Function/Parameter
Setting
Vertical
(sensitivity)
100
mV/div
(channel
key}
Horizontal
{time/div)
50
ns/div
(Setup)
(mode)
repetitive
Display
(mode)
avg
#
of
avg
8
Press
Define
meas_,
then
with
the
softkeys
set
statistics
to
on.
With
the
soft
keys
select
define,
user
defined,
and
measurements.
SS
a
EOE
———
aE
TT
For
valid
statistical
data
In
repetitive
mode,
measurement
specifications
are
valid
with
eight
or
more
acquisitions
averaged.
Statistics
accumulated
before
the
required
number
of
averaged
acquisitions
may
show
the
instrument
to
fail
the
specification.
This
is
particularly
true
for
minimum
and
maximum
in
this
case
since
they
are
set
by
measurements
taken
with
the
fewest
averages.
Ifthe
procedure
above
is
followed
exactly,
the
required
number
of
acquisitions
are
averaged
before
statistics
are
turned
on.
If
however,
Clear
display
is
pressed
after
statistics
are
turned
on,
averaging
and
statistics
are
restarted
simultaneously
and
the
result
is
erroneous
data
collected
from
the
early
averages.
Ifin
doubt
about
the
statistical
data,
after
#Avg
is
complete
press
Define
meas,
select
meas,
)
and
turn
statistics
off
then
on.
This
restarts
the
statistics
without
restarting
averaging
and
the
resultis
valid
data.
Initiate
a
delay
measurement.
Press
(blue)
Aime
1
1
The
A
Time
readings
should
be
25
ns
+601.3
ps;
minimum
24.399
ns
and
maximum
25.601
ns.
Record
the
minimum
and
maximum
readings
in
the
Performance
Test
Record.
Change
the
signal
generator
frequency
to
100
MHz
(10
ns
period).
Set
the
time/div
to
20
ns/div.
Clear
the
statistics.
e
Press
Define
meas
_,
select
measure,
and
turn
statistics
of
f
then
on.
or
e
Press
(blue)
Clr
meas
,thenpress
(blue)
A
‘Time
1
1
The
A
Time
delay
readings
should
be
10
ns
+301
ps:
minimum
9.699
ns
and
maximum
10.301
ns.
Record
the
minimum
and
maximum
readings
in
the
Performance
Test