Chapter
4:
Calibrating
and
Adjusting
To
calibrate
the
firmware
Check
that
default
calibration
factors
are
loaded.
Press
Utility
,then
press
service
menu.
If
default
calibration
factors
are
not
loaded
("D's
in
the
table
on
the
screen),
load
them.
Select
2.
default
cal,
and
press
start
cal
then
continue.
Press
exit
menu
then
self
cal
menu.
Cal
select
0
(0.vertical)
should
be
selected.
Press
channel]
to
select
all,
then
press
start
cal
and
follow
the
instructions
on
the
display.
Calibrating
individual
channels
or
all
These
calibrations
may
be
done
individually,
but
using
all
(when
available},
and
connecting
all
inputs
at
once
avoids
operator
interaction
between
calibrations.
It
also
shortens
the
total
calibration
time
(see
the
"Calibration
Time"
sidebar
on
the
previous
page).
If
the
entire
calibration
procedure
fails
while
running
all,
run
the
calibrations
individually.
[fone
input
is
!oading
the
cal
signal
(an
input
stuck
in
50
Q
for
example),
calibration
will
fail
for
all
inputs.
Individual
calibration
will
isolate
the
failure.
After
completing
vertical
cals,
press
cal
select
toselectcal
select
1
(1.delay).
Press
channel
to
select
channel
1,
then
press
start
cal
and
follow
the
instructions
on
the
display.
When
the
channel
1
calibration
is
complete,
select
the
other
channels
in
turn
and
follow
the
instructions
on
the
display.
Messages
will
be
displayed
as
each
calibration
routine
is
completed
to
indicate
calibration
has
passed
or
failed.
When
the
delay
cal
is
complete,
press
cal
select
toselect
cal
select
2
(2.time
null).
Press
start
cal
and
follow
the
instructions
on
the
display.
Channels
2,
3,
and
4
are
calibrated
against
channel
1
As
each
pair
of
channels
completes
calibration,
select
the
next
and
press
continue.
When
time
null
calibration
is
complete,
press
cal
select
toselectcal
select
3
(3.logic
trigger).
Press
start
cal
and
follow
the
instructions
on
the
display.
Return
to
the
start
of
the
self
cal
menu
(press
cal
select).
Note
that
the
entire
calibration
matrix
is
filled
with
"P,"
and
that
all
tests
have
passed.
After
calibration
has
been
completed,
switch
the
rear-panel
Calibration
switch
to
Protected
(down).
If
calibration
fails
Go
to
chapter
5
Troubleshooting.”