SVC 10-13
The message “PASSED SELF TEST” indicates INET, at least with respect
to the HP 5890A, is performing satisfactorily. If FAILED SELF-TEST is
displayed, a bad cable may be indicated; install a different INET cable and
repeat the test. If FAILED SELF-TEST is displayed again for a second
cable, electronic problems within the HP 5890A are indicated; see Section
19, “Electronic Troubleshooting,” for more information.
The LOOPBACK TEST may be used to check for continuity in an
INET cable;an open cable causestest failure.Verify that the cable
is at fault (rather than the HP 5890A) by testing a second cable.
If an intermittent cable problem is suspected, the test may be re-
peated as necessary while flexing the cable (particularly at its
plugs).Anohmmeter shouldalsobeusedtotestforproblems;itisa
reliable method for testing continuity.
NOTE