6-5
Error Messages
Error Messages in Alphabetical Order
BATTERY LOW! STORE SAVE REGS TO DISK
Error Number
184
The battery protection of the non-volatile CMOS memory is in danger of failing.
If this occurs, all of the instrument state registers stored in CMOS memory will
be lost. Save these states to a disk and refer to the service guide for battery
replacement instructions. See Chapter 8 , “Preset State and Memory Allocation”
for more information about the CMOS memory.
BLOCK INPUT ERROR
Error Number
34
The analyzer did not receive a complete data transmission. This is usually
caused by an interruption of the bus transaction. Clear by pressing the
key or aborting the I/O process at the controller.
BLOCK INPUT LENGTH ERROR
Error Number
35
The length of the header received by the analyzer did not agree with the size of
the internal array block. Refer to the programmer’s guide for instructions on
using analyzer input commands.
CALIBRATION ABORTED
Error Number
74
You have changed the active channel during a calibration so the calibration in
progress was terminated. Make sure the appropriate channel is active and
restart the calibration.
CALIBRATION REQUIRED
Error Number
63
A calibration set could not be found that matched the current stimulus state or
measurement parameter. You will have to perform a new calibration.
CANNOT FORMAT DOS DISKS ON THIS DRIVE
Error Number
185
You have attempted to initialize a floppy disk to DOS format on an external disk
drive that does not support writing to all 80 tracks of the double density and
high density disks. The older single-sided disks had only 66 tracks and some
disk drives were limited to accessing that number of tracks. To format the disk,
either choose another external disk drive or use the analyzer's internal disk
drive.
CANNOT MODIFY FACTORY PRESET
Error Number
199
You have attempted to rename, delete, or otherwise alter the factory preset
state. The factory preset state is permanently stored in CMOS memory and
cannot be altered. If your intent was to create a user preset state, you must
create a new instrument state, save it, and then rename it to "UPRESET". Refer
to Chapter 8 , “Preset State and Memory Allocation” for more detailed
instructions.