1-26
HP 8719/20/22ES Specifications and Characteristics
Instrument Specifications
Test Port Output
Table 1-23 Test Port Output
HP 8719ES/20ES/22ES Test Port Output
Description Specification Supplemental Information
Frequency
Range
HP 8719ES 0.05 to 13.51 GHz
HP 8720ES 0.05 to 20.05 GHz
HP 8722ES 0.05 to 40.05 GHz
Resolution 1 Hz
Stability
Standard ±7.5 ppm, 0˚ to 55˚C, typ.
±3 ppm/year, typ.
Option 1D5 ±0.05 ppm, 0˚ to 55˚C, typ.
±0.5 ppm/year, typ.
CW Accuracy ±10 ppm at 23˚ ±3˚C
Output Power
Level Accuracy
a
a. Absolute power accuracy at a given power level. Includes absolute accuracy and relative
flatness across frequency.
HP 8719ES/20ES
Standard ±2 dB at 0 dBm
Option 007 ±2 dB at +5 dBm
HP 8722ES
Standard ±3 dB at −10 dBm
Option 007 ±3 dB at −5 dBm