2-11
HP 8719/20/22ET Specifications and Characteristics
Instrument Specifications
Test Port Output
Table 2-7 Test Port Output
HP 8719ET/20ET/22ET Test Port Output
Description Specification Supplemental Information
Frequency
Range
HP 8719ET 0.05 to 13.51 GHz
HP 8720ET 0.05 to 20.05 GHz
HP 8722ET 0.05 to 40.05 GHz
Resolution 1 Hz
Stability
Standard ±7.5 ppm, 0° to 55 °C, typ.
±3 ppm/year, typ.
Option 1D5 ±0.05 ppm, 0° to 55°C, typ.
±0.5 ppm/year, typ.
CW Accuracy ± 10 ppm at 23 °C± 3 °C