2-10
HP 8719/20/22ET Specifications and Characteristics
Instrument Specifications
Table 2-6 2.4-mm (50 ) Device Connector Type
HP 8722ET (2.4-mm, 50 )
Description
Specification
50 to 500 MHz
0.5 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 40 GHz
Directivity
a
(dB)
23 23 21 16 15
SourceMatch
a
(dB)
16 20 14 11 10
Load Match
a
(dB)
22 20 20 20 15
Reflection
b
Tracking (dB)
± 3 ± 3 ± 3 ± 3 ± 4
Transmission
a
Tracking (dB)
± 3 ± 3 ± 3 ± 3 ± 3
Tracking
Stability
(Ratio
Measurement)
Magnitude
(dB)
0.02/˚C, typ. 0.03/˚C, typ. 0.03/˚C, typ. 0.04/˚C, typ. 0.06/˚C, typ.
Phase (deg) 0.1/˚C, typ. 0.1/˚C, typ. 0.3/˚C, typ. 0.5/˚C, typ. 1.0/˚C, typ.
Crosstalk
c
(dB)
95 95 95 88 82
a. Does not include effects of cable on test ports.
b. Does not include rolloff below 500 MHz which is typically 18 dB at 100 MHz, and 25 dB
at 50 MHz.
c. Measurement conditions: Normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the
maximum power out or the maximum receiver power.