2-3
HP 8719/20/22ET Specifications and Characteristics
Corrected System Performance (HP 8719/20ET)
Corrected System Performance (HP 8719/20ET)
The specifications in this section apply for measurements made using 10 Hz IF bandwidth,
no averaging, and at an environmental temperature of 23 ±3 °C, with less than 1 °C
deviation from the calibration temperature. Assumes that an isolation calibration was
performed with an averaging factor of 8.
Table 2-1 System Dynamic Range, All Device Connector Types
HP 8719/20ET, All Options, All Cal Kits, All Cables, 10 Hz IF BW
Description Specification Supplemental
Information
System Transmission Dynamic Range
a
a. The System Transmission Dynamic Range is calculated as the difference
between the receiver noise floor and the lesser of either: the source maximum
output or the receiver maximum input. The receiver noise floor is specified as 3
standard deviations above the mean of the linear magnitude noise floor trace
over the specified frequency band.
50 MHz to 840 MHz 102 dB
840 MHz to 8 GHz 104 dB
8 GHz to 20 GHz 104 dB
Table 2-2 System Dynamic Range, All Device Connector Types
HP 8722ET, All Options, All Cal Kits, All Cables, 10 Hz IF BW
Description Specification Supplemental
Information
System Transmission Dynamic Range
a
a. The System Transmission Dynamic Range is calculated as the difference
between the receiver noise floor and the lesser of either: the source maximum
output or the receiver maximum input. The receiver noise floor is specified as 3
standard deviations above the mean of the linear magnitude noise floor trace
over the specified frequency band.
50 MHz to 840 MHz 98 dB
840 MHz to 8 GHz 102 dB
8 GHz to 20 GHz 100 dB
20 GHz to 40 GHz 89 dB