2-4
HP 8719/20/22ET Specifications and Characteristics
Corrected System Performance (HP 8719/20ET)
Table 2-3 3.5-mm (50 ) Device Connector Type
Network Analyzer: HP 8719/20ET Standard or Option 004 Attenuator
Calibration Kit: HP 85052B (3.5-mm, 50 ) Cal Kit
Cables: HP 85131E Cables
Calibration: One-Port, Response, or Enhanced Response
IF BW = 10 Hz, Avg = off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Specification
50 to 500 MHz 0.5 to 2 GHz 2 to 8 GHz 8 to 20 GHz
Reflection Measurements
a
Directivity (dB) 48 48 44 44
Source Match (dB) 40 40 33 31
Load Match (dB)
One-Port Cal 22 22 22 17
Tracking
Magnitude (dB) ±(0.006 + .02/° C) ±(0.006 + .03/° C) ±(0.006 + .03/° C) ±(0.008 + .04/° C)
Phase (deg) ±(0.040 + .1/°C) ±(0.040 + .1/°C) ±(0.040 + .3/°C) ±(0.053 + 0.5/°C)
Transmission Measurements
b
Source Match (dB)
Enhanced
Response Cal
40 40 33 31
Response Only Cal 16 20 14 11
Load Match (dB) 22 22 22 17
Tracking
Enhanced
Response Cal
Magnitude (dB) ±(0.014 + .02/° C) ±(0.012 + .03/° C) ±(0.027 + .03/° C) ±(0.050 + .04/° C)
Phase (deg) ±(0.092 + .1/°C) ±(0.079 + .1/°C) ±(0.178 + .3/°C) ±(0.33 + 0.5/°C)
Response Only Cal
Magnitude (dB) ±(0.109 + .02/° C) ±(0.069 + .03/° C) ±(0.137 + .03/° C) ±(0.339 + .04/° C)
Phase (deg) ±(0.719 + .1/°C) ±(0.455 + .1/°C) ±(0.904 + .3/°C) ±(2.237 + 0.5/°C)
Uncertainty graphs are on the following page.
a. One-port or enhanced response calibration.
b. Enhanced response or response only calibration.