1-25
HP 8719/20/22ES Specifications and Characteristics
Instrument Specifications
Table 1-22 2.4-mm Device Connector Type
HP 8722ES (2.4-mm, 50 )
Description Characteristic
50 to
500 MHz
0.5 to
2 GHz
2 to
8 GHz
8 to
20 GHz
20 to
40 GHz
Directivity
a
23 23 21 16 15
Source Match
a
Standard/
Option 400 (dB) 17 17 12 11 7
Option 007 or
Option 085 (dB) 17 17 15 11 8
Load Match
a
Standard/
Option 400 (dB)
18 18 15 12 10
Option 007 or
Option 085 (dB)
21 21 17 13 10
Reflection
Tracking
b
(dB)
±3 ±3 ±3 ±3 ±6
Transmission
Tracking
b
±3 ±3 ±3 ±3 ±6
Tracking Stability
(Ratio
Measurement)
Magnitude (dB) 0.02/°C, typ. 0.03/°C, typ. 0.03/°C, typ. .04 dB/˚C, typ. .06 dB/˚C, typ.
Phase (dB) 0.1/°C, typ. 0.1/°C, typ. 0.3/°C, typ. 0.5 ˚/˚C, typ. 1.0 ˚/˚C, typ.
Crosstalk
c
60
d
85 85 82 72
a. Does not include the effect of the cable set on the test ports.
b. Excludes rolloff below 500 MHz, which is typically 18 dB at 100 MHz, and 25 dB at
50 MHz.
c. Measurement conditions: Normalized to a thru, measured with two shorts, 10 Hz IF
bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of the
maximum power out or the maximum receiver power.
d. Limited by noise floor.