Error Messages
Error Messages in Alphabetical Order
NO MEMORY AVAILABLE FOR INTERPOLATION
Error Number
123
You cannot perform interpolated error correction due to insufficient memory.
NO MEMORY AVAILABLE FOR SEQUENCING
Error Number
126
You cannot modify the sequence due to insufficient memory.
NO SPACE FOR NEW CAL. CLEAR REGISTERS
Error Number
70
You cannot store a calibration set due to insufficient memory. You can free more
memory by clearing a saved instrument state from an internal register (which
may also delete an associated calibration set, if all the instrument states using
the calibration kit have been deleted). You can store the saved instrument state
and calibration set to a disk before clearing them. After deleting the instrument
states, press to run the memory packer.
NOT ALLOWED DURING POWER METER CAL
Error Number
198
When the analyzer is performing a power meter calibration, the HP-IB bus is
unavailable for other functions such as printing or plotting.
NOT ENOUGH SPACE ON DISK FOR STORE
Error Number
44
The store operation will overflow the available disk space. Insert a new disk or
purge files to create free disk space.
NO VALID MEMORY TRACE
Error Number
54
If you are going to display or otherwise use a memory trace, you must first store
a data trace to memory.
NO VALID STATE IN REGISTER
Error Number
55
You have requested the analyzer, over HP-IB (or by sequencing), to load an
instrument state from an empty internal register.
OK TO ALTER CORRECTION CONSTANTS?
Error Number
186
ONLY LETTERS AND NUMBERS ARE ALLOWED
Error Number
43
You can only use alpha-numeric characters (and underscores) in disk file titles
or internal save register titles. Other symbols are not allowed, except for the
"underscore" symbol.