1-30
HP 8719/20/22ES Specifications and Characteristics
Instrument Specifications
Table 1-27 Test Port Output
HP 8719ES/20ES Test Port Output
Description Specification Supplemental Information
Linearity
a
Test Reference Powers:
Standard HP 8719/20ES: 5 dBm
HP 8719/20ES Option 007: 0 dBm
5 dB from reference ±0.35 dB
+5 dB from reference ±0.35 dB
10 dB from reference ±0.6 dB
+10 dB from reference ±1.0 dB
Impedance
Standard 50 , nominal
Attenuator Accuracy
b
0 dB reference; at 50 MHz
5 dB ±0.6 dB, char.
10 dB ±0.9 dB, char.
15 dB ±1.25 dB, char.
20 dB ±1.5dB, char.
25 dB ±2.0 dB, char.
30 dB ±2.5 dB, char.
35 dB ±2.8 dB, char.
40 dB ±3.0 dB, char.
45 dB ±3.1 dB, char.
50 dB ±3.2 dB, char.
55 dB ±3.2 dB, char.
a. Change in source output power for a given change in source power setting at any given
frequency.
b. The accuracy, relative to the 0 dB setting, of each setting of an attenuator, at a given
frequency.