HP 8719/20/22ET Specifications and Characteristics
Corrected System Performance (HP 8722ET)
Table 2-4 2.4-mm (50 ) Device Connector Type
Network Analyzer: HP 8722ET Standard or Option 004 Attenuator
Calibration Kit: HP 85056A (2.4-mm, 50 )
Cables: HP 85133E Cables
Calibration: One-Port, Response, or Enhanced Response
IF BW = 10 Hz, Avg = off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Specification
0.05 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 40 GHz
Reflection Measurements
a
Directivity (dB) 42 42 42 38
Source Match (dB) 41 38 38 33
Load Match (dB)
One-Port Cal 22 20 20 15
Tracking
Magnitude (dB) ±(0.005 + .03/°C) ±(0.010 + .03/°C) ±(0.010 + .04/°C) ±(0.021 + .06/ °C)
Phase (deg) ±(0.033 + .1/°C) ±(0.066 + .3/°C) ±(0.066 + .5/°C) ±(0.139 + 1.0/°C)
Transmission Measurements
b
Source Match (dB)
Enhanced
Response Cal
41 38 38 33
Response Only
Cal
16 14 11 10
Load Match (dB) 22 20 20 15
Tracking
Enhanced
Response Cal
Magnitude (dB) ±(0.018 + .03/°C) ±(0.026 + .03/ °C) ±(0.031 + .04/°C) ±(0.069 + .06/°C)
Phase (deg) ±(0.119 + .1/°C) ±(0.172 + .3/°C) ±(0.205 + .5/°C) ±(0.455 + 1.0/°C)
Response Only
Cal
Magnitude (dB) ±(0.109 + .03/°C) ±(0.172 + .03/°C) ±(0.241 + .04/°C) ±(0.475 + .06/°C)
Phase (deg) ±(0.719 + .01/°C) ±(1.135 + .3/°C) ±(1.59 + .5/°C) ±(3.135 + 1.0/°C)
Uncertainty graphs are on the following page.
a. One-port or enhanced response calibration.
b. Enhanced response or response only calibration.