2-33
HP 8719/20/22ET Specifications and Characteristics
Instrument Specifications
Table 2-25 General Information
HP 8719/20/22ET General Information
Description Specification Supplemental Information
General Environmental
RFI/EMI Susceptibility Defined by CISPR Pub. 11 and
FCC Class B standards.
ETD Minimize using static-safe
work procedures and an
antistatic bench mat
(HP part number 9300-0797).
Dust Minimize for optimum
reliability.
Operating Environment
Temperature 0° C to +55° C Instrument powers up, phase
locks, and displays no error
messages within this
temperature range.
Error-corrected temperature
range
see system specifications
Humidity 5% to 95% at +40 °C
(non-condensing)
Altitude 0 to 4.5 km (15,000 ft)
Storage Conditions
Temperature 40 °C to +70 °C
Humidity 0% to 95% RH at +65 °C
(non-condensing)
Altitude 0 to 15.24 km (50,000 ft)
Cabinet Dimensions
Height x Width x Depth 222 x 425 x 457 mm, nom.
(8.75 x 16.75 x 18 in, nom.)
Cabinet dimensions exclude
front and rear protrusions.
Weight
Shipping 41kg (90 lb), nom.
Net 27 kg (60 lb), nom.
Internal Memory - Data Retention Time with 3 V, 1.2 Ah Battery
a
70° C 250 days (0.68 year), char.
40° C 1244 days (3.4 years), char.
25° C 10 years, char.
a. Analyzer power is switched off.