EasyManuals Logo

HP EPM-441A User Manual

HP EPM-441A
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
HP4402.book : chapter4.fm 3 Tue Apr 22 15:20:23 1997
Error Messages
Introduction
HP EPM-441A User’s Guide 4-3
Errors are retrieved in a first in first out (FIFO) order. If more than 30
errors occur, the error queue overflows and the last error in the queue is
replaced with error -350, “Queue Overflow”. Any time the queue
overflows the most recent error is discarded.
Once the errors are read they are removed from the error queue. This
opens a position at the end of the queue for a new error message, if one is
subsequently detected. When all errors have been read from the queue,
further error queries return +0, “No errors”.
To delete all the errors in the queue from the front panel press:
, , then .
To delete all the errors in the queue remotely use:
the *CLS (clear status) command.
The error queue is also cleared when the instrument power has been
switched off.
System
Inputs
More
Error List Clear Errors