Outer Dataset® Arm (Calibration Tool, 221-672-1, Required)
With the hood in the raised position, verify that the outer arm is in the “home” position
and that the arm and hood are not moving. Tap the foot pedal once or press "Enter.”
Place the calibration tool on the hub using the hub mounting slot as shown.
2
3
1
Place the outer Dataset® arm ball in hole position “1.” Tap the foot pedal once or
press “Enter.”
Place the outer Dataset® arm ball in hole position “2.” Tap the foot pedal once or
press “Enter.”
Place the outer Dataset® arm ball in hole position “3.” Tap the foot pedal once or
press “Enter.”
Dataset® arm calibration is complete. “RDY” is displayed near the inner arm graphic
to indicate that the inner Dataset® calibration passed. “RDY --]” is displayed on the
weight digits to indicate that the outer Dataset® calibration passed.
Enabling/Disabling Inflation Station and Changing Units
Press the “Setup/Cal” button twice. “Setup” is shown on the left and right digits. Press
the “Enter” button to begin setup.
Press “Enter” until “INF” is shown on the left weight digits.
Press the “Next” button to select pressure units (kPa, PSI, or Bar) . “OFF” in the right
weight digits indicates that Inflation Station is disabled.
Press “Enter” to store selection, then press “Setup/Cal” button to exit setup.
Inflation Station Calibration
Verify that Inflation Station is enabled (see previous section).
Press the “Setup/Cal” button once. “Cal” is displayed.
Press the “Next” button twice. “Inf Sta” is displayed on left and right weights digits.
Press “Enter” to begin calibration. “-0- PSI” is displayed (Or other units, refer to
“Enabling/Disabling Inflation Station and Changing Units,” page 11).
Store the zero pressure reading by pressing “Enter” with the inflation hose in the
home storage position.
“OPT Ref” is displayed. This is an optional step to calibrate to a reference gauge.
Measure a tire’s pressure using the pressure gauge that you wish to use as a
reference. Enter the measured pressure by turning the left control knob (distance
dimension).
DSP9600/9100 Service Manual Calibration Procedures 11