Status bits for the Balancer Spin Calibration
If calibration passed, the bits will be “111111 111111”.
The bits are described below.
Bit 1 - This bit verifies that spin 1 (bare shaft) left force transducer
readings are within tolerance. If this bit fails, the cal weight may
have been installed before spin 1.
Bit 2 - This bit verifies that spin 1 (bare shaft) right force transducer
readings are within tolerance. If this bit fails, the cal weight may
have been installed before spin 1.
Bit 3 - This bit verifies that spin 2 left force transducer readings are
within tolerance. If this bit fails, the cal weight may not have
been installed, or was installed on the wrong side.
Bit 4 - This bit verifies that spin 2 right force transducer readings are
within tolerance. If this bit fails, the cal weight may not have
been installed, or was installed on the wrong side.
Bit 5 - This bit verifies that the cal weight was moved to the opposite
side between spins 2 and 3.
Bit 6 - This bit verifies that the cal weight was in the same hole for spins
2 and 3.
NOTE: Press “Next” to show remaining bits (1b). The first bit on the
right will be bit 7.
Bit 7 - This bit verifies that the left force transducer temperature reading
was within tolerance. If this bit fails, the transducer or DSP
control board is defective.
Bit 8 - This bit verifies that the right force transducer temperature
reading was within tolerance. If this bit fails, the transducer or
DSP control board is defective.
Bit 9 - This bit verifies that the “4x” gain stage for the left force
transducer is operating properly. If this bit fails, the transducer or
DSP control board is defective.
Bit 10 - This bit verifies that the “4x” gain stage for the right force
transducer is operating properly. If this bit fails, the transducer or
DSP control board is defective.
Bit 11 - This bit verifies that the “2x” gain stage for the left force
transducer is operating properly. If this bit fails, the transducer or
DSP control board is defective.
Bit 12- This bit verifies that the “2x” gain stage for the right force
transducer is operating properly. If this bit fails, the transducer or
DSP control board is defective.
NOTE: Turn the “D” knob to advance to next diagnostic variable.
DSP9600/9100 Service Manual Calibration Results • 13