User Manual
HV Diagnostics Page 8 Rev 9.0
DANGER
Authorized Personnel Only
The test area must be secured to keep non-qualified personnel off the
premises!
Signs must warn all persons of the high voltage test area.
Only qualified electrical technicians should have access to the test
area.
Other persons must be accompanied by qualified electrical
technicians and must be informed of the risks involved.
WARNING
Radiation Hazard
Testing vacuum bottles, above their voltage rating, with DC can produce
dangerous X-rays.
NOTICE
Equipment Handling
DUT must have clean connections.
Test instruments must only be repaired or modified by authorized
HV Diagnostics’ personnel.
NOTICE
If required according to local safety regulations
Wear high voltage gloves when handling high voltage cables and
equipment.