Hydroid Series User Manual
27 04/19
TROUBLESHOOTING
SYMPTOMS
POSSIBLE CAUSES
CORRECTIVE ACTION
Low Inlet Pressure
Low supply pressure
Increase inlet pressure
Pre-filter is plugged
Change pre-filter
Solenoid valve malfunction
Inspect / replace solenoid valve and/or coil
Leaks
Fix any visible leaks
Low PRODUCT
(purified water) Flow
Cold feedwater
See temperature correction sheet
Low operating pressure
See low inlet pressure
Fouled or scaled membrane
Replace membranes
Pre-filter is plugged
Change pre-filter
High PRODUCT
(purified water) Flow
Damaged or oxidized
membrane
Replace membranes
Exceeding maximum
feedwater temperature
See temperature correction sheet
Poor PRODUCT
(purified water) Quality
Low operating pressure
See low inlet pressure
Damaged or oxidized
membrane
Replace membranes
System Inoperative
Fuse
If water quality display not lit, check fuse F2 below
transformer
Wiring to controller
If fuse is ok, check power terminals to L1 and L2 with
A voltmeter. If no power, rewiring may be required.
Defective controller board
If power to terminals is OK, the board may be
defective and should be replaced.
Inlet Valve Will Not
Operate
Controller is off
Turn controller on
Fuse
If shutdown conditions are active and inlet LED, L2 is
lit, check fuse F1.
Defective controller board
If shutdown conditions are active and inlet LED L2 is
NOT lit, board may need to be replaced.
Wiring to valve
Use voltmeter to check inlet terminals for power to
controller board. If board has power, check valve and
valve wiring.
RO Pump Will Not
Operate
Controller is off
Turn controller on
Fuse
If shutdown conditions are active and RO LED L1 is
lit, check fuse F1.
Defective controller board
If shutdown conditions are active and inlet LED L1 is
NOT lit, board may need to be replaced.
Wiring to pump
Use voltmeter to check inlet terminals for power to
controller board. If board has power, pump motor and
motor wiring.