Manufacturer’s EMC guidelines
43
Electromagnetic immunity
The Tel-O-Graph
®
GSM is intended to be operated in the electromagnetic environment specified below. Only use
the Tel-O-Graph
®
GSM in such an environment.
Measurement of
interference emissions
Electromagnetic environment –
guidelines
Electrostatic discharge
(ESD) in accordance with
IEC 61000-4-2
+/- 6kV contact
discharge
+/- 8kV air
discharge
+/- 6kV contact
discharge
+/- 8kV air discharge
Floors should consist of wood or cement
or ceramic tiles. If the floor consists of
synthetic materials, relative humidity must
be at least 30%.
Magnetic field at supply
frequency (50/60 Hz) in
accordance with
IEC 61000-4-8
Magnet fields at mains frequency should
match the typical values found in
business and hospital environments.
HF radiated disturbances
in accordance with
Portable and mobile wireless receiving
and transmission equipment should be