Appendix 2
Diagrams of the sterilization programs STE-18-T
The time required for sterilizer to be ready for routine use after the power is switched is less than 5
minutes.
The max. Temperature of the 134°C sterilization cycle is 137°C
The max. Temperature of the 121°C sterilization cycle is 124°C
Programs
Temperature (°C)
Pressure (kPa)
Holding time (min)
Total time (min)
Type
Max load (kg)
Max load per tray
(kg)
Quick S 134 210 4 11-15 Unwrapped solid material 1.00 0.8
Standard B 134 210 4 30-49
Unwrapped solid material 4.50 1.50
Single-wrapped solid or hollow
material
4.00 1.20
Soft B 121 110 20 38~54
Unwrapped solid material
4.50 1.50
Single-wrapped solid or hollow
material
4.00 1.20
Prion 134 210 18 37-52
Unwrapped porous material 1.25 0.40
Single-wrapped porous material 1.10 0.30
Dual-wrapped porous material 0.75 0.25
Single-wrapped hollow material 4.00 1.25
Dual-wrapped solid and hollow
material
2.00 0.60
Drying
(optional)
― ― ― 1-20 ― ― ―
B&D test 134 210 3.5 28-35 ― ― ―
Helix test 134 210 3.5 28-35 ― ― ―
Vacuum test ― ― ― 20-25 ― ― ―