Page 16
*Press the Blue button to enter AC Voltage function
AC Voltage Test: Model 61-481 and 61-482
Range Source Frequency Reading
400mV 370.0mV 50Hz 361.8 to 378.4
400mV 370.0mV 60Hz 361.8 to 378.4
4V 3.700V 50Hz 3.647 to 3.753
4V 3.700V 500Hz 3.647 to 3.753
40V 37.00V 50Hz 36.47 to 37.53
40V 37.00V 1KHz 36.47 to 37.53
400V 370.0V 50Hz 364.7 to 375.3
400V 370.0V 1KHz 364.7 to 375.3
750V 700V 50Hz 686 to 714
750V 700V 1KHz 686 to 714
DC Current Test: Model 61-480
Range Source Reading
30mA 20.00mA 19.68 to 20.32
300mA 200.0mA 196.8 to 203.2
1A .90A .86 to .94
10A 9.0A 8.79 to 9.21
DC Current Test: Model 61-481 and 61-482
Range Source Reading
40mA 37.00mA 36.76 to 37.24
400mA 370.0mA 367.3 to 372.7
1A .90A 0.88 to .92
10A 9.0A 8.88 to 9.12
*Press the Blue button to enter AC Current function
AC Current Test: Model 61-480
Range Source Frequency Reading
30mA 20.00mA 50Hz 19.55 to 20.45
30mA 20.00mA 1KHz 19.55 to 20.45
300mA 200.0mA 50Hz 195.5 to 204.5
300mA 200.0mA 1KHz 195.5 to 204.5
1A .90A 50Hz 0.83 to .97
1A .90A 1KHz 0.83 to .97
10A 9.0A 50Hz 8.73 to 9.27
10A 9.0A 1KHz 8.73 to 9.27