*Press the Blue button to enter AC Current function
AC Current Test: Model 61-481 and 61-482
Range Source Frequency Reading
40mA 37.00mA 50Hz 36.21 to 37.79
40mA 37.00mA 1KHz 36.21 to 37.79
400mA 370.0mA 50Hz 362.1 to 377.9
400mA 370.0mA 1KHz 362.1 to 377.9
1A .90A 50Hz 0.83 to .97
1A .90A 1KHz 0.83 to .97
10A 9.0A 50Hz 8.73 to 9.27
10A 9.0A 1KHz 8.73 to 9.27
Resistance Test: Model 61-480
Range Source Reading
300 180
177.8 to 182.2
3K 1.800K
1.785 to 1.815
30K 18.00K
17.85 to 18.15
300K 180.0K
178.5 to 181.5
3M 1.800M
1.779 to 1.821
30M 29.00M
28.37 to 29.63
*Lead resistance on the 300 range is not included in error.
Resistance Test: Model 61-481 and 61-482
Range Source Reading
400 360
357.2 to 362.8
4K 3.600K
3.583 to 3.617
40K 36.00K
35.83 to 36.17
400K 360.0K
358.3 to 361.7
4M 3.600M
3.57 to 3.62
40M 39.00M
38.37 to 39.63
*Lead resistance on the 400 range is not included in error.
*Press the Blue button to test
function
Continuity Beep Test: All Models
Range Source Test
20
.Beeper on
50
Beeper off
Buzzer will Alarm if resistance is less than approximately 30
*Press the Blue button to test
function
Diode Test: All Models
Range Source Reading
500mV DC Approx. .485 - .515
Form number TM61480-1-2 Rev 5 July 2002