Page 7
Peak Hold:
Peak hold table for Model 61-482 only
Function Range Accuracy
400mV
Unspecified
4V
± (1.5% + 300 digits)
40V
400V
DCV
1000V
± (1.5% + 60 digits)
400mV
Unspecified
4V
± (1.5% + 300 digits)
40V
400V
ACV
1000V
± (1.5% + 60 digits)
Function Range Accuracy
40mA
400mA
± (3.0% + 60 digits)
DCA
10 A
± (1.5% + 60 digits)
40mA
400mA
± (3.0% + 60 digits)
ACA
10 A
± (1.5% + 60 digits)
Resistance (Ohms) for Model 61-480
Range Resolution Accuracy
Over Voltage
Protection
300 Ω .1 Ω ± (1.0% + 4 digits)
3K Ω 1 Ω
30K Ω 10 Ω ± (0.4% + 3 digit)
600V rms
300K Ω .1K Ω
3M Ω 1K Ω ± (0.6% + 3 digit)
30M Ω 10K Ω ± (1.5% +5 digit)
Open Circuit Voltage: 1.5V approx.
Resistance (Ohms) for Model 61-481-and 61-482
Range Resolution Accuracy
Over Voltage
Protection
400 Ω .1 Ω ± (0.7% + 3 digits)
4K Ω 1 Ω
40K Ω 10 Ω ± (0.4% + 3 digit)
600V rms
400K Ω .1K Ω
4M Ω 1K Ω ± (0.6% + 3 digit)
40M Ω 10K Ω ± (1.5% +5 digit)
Open Circuit Voltage: 1.3V approx.
Form number TM61480-1-2 Rev 5 July 2002