Operating modes
Operating mode A
Switch on the device at the rocker switch (B).
The heating function is switched off.
Safe Temp (safety temperature) will appear for 5 secs (fig. 4).
The target value is set to 20 °C.
The target value setting can be adjusted.
The controller tempering medium is set to water.
The heating function must be restarted following a power failure.
The following settings are possible/adjustable:
- the target value,
- the safety circuit,
- the controller tempering medium.
When the device is switched on at the mains switch, A appears
(fig. 11).
Factory setting: Mode A
Operating mode B
Switch on the device at the rocker switch (B).
The heating function is switched off.
Safe Temp (safety temperature) will appear for 5 secs (fig. 4).
The target value is set to 20 °C or the most recent temperatu-
re setting.
The controller tempering medium is set to water or the most
recent medium setting.
The heating function must be restarted following a power failure.
The following settings are possible/adjustable:
- the target value setting,
- the controller tempering medium.
The safety circuit is set to the most recently set value and can-
not be adjusted.
When the device is switched on at the mains switch, B appears.
Operating mode C
Switch on the device at the rocker switch (B).
The heating function is switched on or off depending on the last
selected setting.
Safe Temp (safety temperature) will appear for 5 secs (fig. 4).
The last target value set in mode B applies.
The setting controller tempering medium set before the device
was last switched off in mode B applies.
The following settings cannot be adjusted:
- the target value,
- the safety circuit,
- the controller tempering medium.
When the device is switched on at the mains switch, C appears.
Switching operating modes
Switch the device off at the mains switch.
Press and hold knob (E) and switch the device on at the mains
switch. Release the knob (E) after approx. 2 seconds.
Sequence A, B, C, A etc.
The operating mode setting appears on the display when star-
ting up the device (see fig. 11).
Safety circuit
The adjustable safety circuit prevents the temperature of the hea-
ting bath from getting too high as a result of
- a controller error,
- the "Temperature" rotating knob having been accidentally turned.
Once the safety temperature has been reached, the device
switches off permanently.
19