23
Any malfunctions during operation will be identified by an error message on the display (H), see fig. 6.
Proceed as follows in such cases:
Switch off device using the main switch (B)
Carry out corrective measures
Restart device
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Error message HB 10 digital/control
Error message Cause Effect Correction
E 2 - In remote operation (PC), no communication Heating off - Switch off the device
between RV 10 and HB 10 - Clean the IR interface or
- IR connection to RV 10 disconnected remove objects from around the IR interface
- Switch on the device
E 3 - Temperature inside the device is too high Heating off - Switch off the device
- Let the device cool down
- Switch on the device
E 9 - Fault in storing the safety circuit set temperatures- Heating off - Switch off the device
- Memory chip (EPROM) defective - Let the device cool down
- Switch on the device
E 21 - Safety relay does not open Heating off - Switch off the device and switch on again
E 23 - Adjustable safety circuit defective Heating off - Switch off the device and switch on again
- Allow the device to cool down
- Switch on the device
E 24 - Set safety temperature exceeded Heating off - Switch off the device and switch on again
- Allow the device to cool down
- Switch on the device again
- Check the safety temperature setting
E 25 - The heater control circuit switch (TRIAC) is faulty Heating off - Switch off the device and switch on again
The heater or the supply line is disconnected
E 26 - Dry running Heating off - Let the device cool down
- Fill up the tempering medium
- Switch off the device and switch on again
E 27 - Error in calibration Heating off - Switch off the device and switch on again
E 28 - Breakage of the controller sensor Heating off - Switch off the device and switch on again
E 29 - Short-circuit in the safety sensor Heating off - Switch off the device and switch on again
E 30 - Short-circuit in the controller sensor Heating off - Switch off the device and switch on again
3946600_HB10bdc_032009:3345300_VG3_042005_kvmb_1 28.04.2009 9:23 Uhr Seite 23