Any malfunctions during operation will be identified by lighting up the red signal lamp “Error”.
Proceed as follows in such cases:
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Error message HB 10 basic
Error message Cause Effect Correction
”Error” signal lamp - Safety circuit temperature set too low Device off - Let device cool down
lights up - Dry run - Reset
- Controller set higher than safety circuit - Check settings and correct if necessary
- Top up tempering medium
”Error” signal lamp - Faulty controller Device defect - Contact the service department.
lights up again after
reset, also in case of
max. adjusted
safety temperature
limit